2400 Series SourceMeter
®
User’s Manual SCPI Command Reference 18-41
Description This command is used to define the 3-bit or 4-bit output pattern for the
Digital I/O Port when there are no failures. Note that the output value
can be specified in binary, octal, decimal, or hexadecimal format. Use
the table provided in the “Description” for the :SOURce command to
determine the parameter value for the desired decimal digital output
pattern.
The SourceMeter can be configured to place the defined pass bit pattern
on the digital output immediately when the pass condition occurs, or it
can wait until all testing on a device package is completed (operation
leaves trigger layer). See “Composite testing,” page 18-40 and
“BCONtrol <name>,” page 18-43, for details.
For the sorting mode, this command defines the 3-bit or 4-bit output
pass pattern for the Limit 1 test (compliance) when Limits 2, 3, 5-12 are
disabled.
FAIL:SOURce2 <NRf> | <NDN>
:CALCulate2:CLIMits:FAIL:SOURce2 <NRf> | <NDN> Specify fail pattern
Parameters <NRf> = 0 to 7 (3-bit) Decimal value
0 to 15 (4-bit) Decimal value
<NDN> = 0 to #b111 (3-bit) Binary value
0 to #b1111 (4-bit) Binary value
0 to #q7 (3-bit) Octal value
0 to #q17 (4-bit) Octal value
0 to #h7 (3-bit) Hexadecimal value
0 to #hF (4-bit) Hexadecimal value
Query :SOURce2? Query programmed source value
Description For the sorting mode, this command is used to define the 3-bit or 4-bit
output pattern for the Digital I/O Port when there are failures. Note that
the output value can be specified using binary, octal, decimal, or hexa-
decimal format. Use the table provided in the “Description” for the
:SOURce command to determine the decimal parameter value for the
desired digital output pattern.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com