3-6 Basic Source-Measure Operation 2400 Series SourceMeter
®
User’s Manual
Compliance limit
When sourcing voltage, the SourceMeter can be set to limit current. Conversely, when
sourcing current, the SourceMeter can be set to limit voltage. The SourceMeter output will
not exceed the compliance limit. Table 3-2 summarizes compliance limits according to
range. See Section 6 for more details on compliance limit.
Table 3-2
Compliance limits
2400 2410 2420
Measure
range
Maximum
compliance
value
Measure
range
Maximum
compliance
value
Measure
range
Maximum
compliance
value
200mV
2V
20V
200V
±210mV
±2.1V
±21V
±210V
200mV
2V
20V
1000V
±210mV
±2.1V
±21V
±1.1kV
200mV
2V
20V
60V
±210mV
±2.1V
±21V
±63V
1µA
10µA
100µA
1mA
10mA
100mA
1A
±1.05µA
±10.5µA
±105µA
±1.05mA
±10.5mA
±105mA
±1.05A
1µA
10µA
100µA
1mA
20mA
100mA
1A
±1.05µA
±10.5µA
±105µA
±1.05mA
±21mA
±105mA
±1.05A
10µA
100µA
1mA
10mA
100mA
1A
3A
±10.5µA
±105µA
±1.05mA
±10.5mA
±105mA
±1.05A
±3.15A
2425/2430 2440
Measure
range
Maximum
compliance
value
Measure
range
Maximum
compliance
value
200mV
2V
20V
100V
±210mV
±2.1V
±21V
±105V
200mV
2V
10V
40V
±210mV
±2.1V
±10.5V
±42V
10µA
100µA
1mA
10mA
100mA
1A
3A/10A
±10.5µA
±105µA
±1.05mA
±10.5mA
±105mA
±1.05A
*
10µA
100µA
1mA
10mA
100mA
1A
5A
±10.5µA
±105µA
±1.05mA
±10.5mA
±105mA
±1.05A
±5.25A
* ±3.15A (2425 and 2430 DC mode)
±10.5A (Pulse mode; 2430 only)