F-4 Contact Check Function 2400 Series SourceMeter
®
User’s Manual
Operation
DUT connections
The contact check function can be used to verify satisfactory contact resistance to the
DUT at both the front and rear panels. If you do not wish to perform contact check using
the GUARD/GUARD SENSE jacks (on the rear panel), they must be connected for proper
operation. The SourceMeter must be placed into remote sense (4-wire) mode for contact
check to function. If in 2-wire mode, contact check can be enabled but will not be
performed.
All three source/sense pairs are tested when contact check is enabled. If a source/sense
pair is not used in a particular test, it must have a shorting plug installed. Other than this
requirement, normal connections to the DUT should be made as shown in Figure F-3.
Figure F-3
Contact check DUT connections
WARNING: NO INTERNAL OPERATOR SERVICA
CAUTION: FOR CONTINUED PROTECTION AGAINST FI
250V
PEAK
250V
PEAK
HI
LO
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
250V
PEAK
5V
PEAK
5V
PEAK
5V
PK
RS232
V, ,
GUARD
4-WIRE
SENSE
INPUT/
OUTPUT
GUARD
SENSE
Contact
Being Tested
Note: Jumper V, , GUARD, and
GUARD SENSE when using front
panel terminals for contact check.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2410

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2410 and is the answer not in the manual?

Keithley SourceMeter 2410 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2410
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals