11-12 Triggering 2400 Series SourceMeter
®
User’s Manual
Figure 11-3
Remote trigger model
Arm-In
Event
IMMediate
BUS
TIMer
MANual
TLINk
NSTest
PSTest
BSTest
Arm Event
Detector
Source Event
Detector
Trigger Delay
SOURCE
Action
Delay Event
Detector
DELAY**
Action
Measure
Event
Detector
MEASURE
Action
No
Yes
Idle
Another
Arm
?
Another
Trigger
?
Yes
No
Arm
Layer
Trigger
Layer
= GPIB Default
= Output Trigger
Trigger-In
Source
1
1
0.0 sec
0.001 sec
No
See Note
INITiate
?
Yes
SOURce
ARM
:DIRection
ACCeptor
ARM :SOURce
TRIGger
:DIRection
SOURce
ACCeptor
TRIGger:INPut
SOURce
TRIGger:DELay
<n>
SOURce
TRIGger:OUTPut*
SOURce:DELay
<n>|AUTO
DELay
TRIGger:OUTPut*
TRIGger:OUTPut*
SENSe
TRIGger:INPut
TRIGger:INPut
SENSe
DELay
TRIGger:COUNt
<n>
ARM:OUTPut
NONE|TEXit
ARM:COUNt
<n>|INF
Note:
The following commands
place the SourceMeter into
idle: DCL, SDC, ABORt,
*RST, SYSTem:PREset and
*RCL.
In :SYSTem:RCMode MULTiple, the soak time programmed with
:SOURce[1]:SOAK takes the place of the delay time only during the
first SDM cycle after the initial sweep trigger. See Section 17 for details.
*
TRIGger:SOURce
IMMediate
TLINk
GPIB default parameter for TRIGGER:OUTPut is NONE.
**
ARM:OUTPut NONE|TENTer
✛
✛
✛
✛
✛
✛
✛
✛
✛
✛
✛
✛