2400 Series SourceMeter
®
User’s Manual SCPI Command Reference 18-65
[:UPPer] <n>
[:SENSe[1]]:CURRent[:DC]:RANGe[:UPPer] <n>|UP|DOWN Select range for amps
[:SENSe[1]]:VOLTage[:DC]:RANGe[:UPPer] <n>|UP|DOWN Select range for volts
[:SENSe[1]]:RESistance:RANGe[:UPPer] <n>|UP|DOWN Select range for ohms
Parameters 2400
<n> = -1.05 to 1.05 Expected reading in amps
-210 to 210 Expected reading in volts
0 to 2.1e8 Expected reading in ohms
DEFault 1.05e-4 (amps), 21 (volts), 2.1e5
(ohms)
MINimum -1.05 (amps), -210 (volts), 0 (ohms)
MAXimum 1.05 (amps), 210 (volts), 2.1e8 (ohms)
UP Select next higher measurement range
DOWN Select next lower measurement range
2410
<n> = -1.05 to 1.05 Expected reading in amps
-1100 to 1100 Expected reading in volts
0 to 2.1e8 Expected reading in ohms
DEFault 1.05e-4 (amps), 21 (volts), 2.1e5
(ohms)
MINimum -1.05 (amps), -1100 (volts), 0 (ohms)
MAXimum 1.05 (amps), 1100 (volts), 2.1e8 (ohms)
UP Select next higher measurement range
DOWN Select next lower measurement range
2420
<n> = -3.15 to 3.15 Expected reading in amps
-63 to 63 Expected reading in volts
0 to 2.1e7 Expected reading in ohms
DEFault 1.05e-4 (amps), 21 (volts), 2.1e5
(ohms)
MINimum -3.15 (amps), -63 (volts), 0 (ohms)
MAXimum 3.15 (amps), 63 (volts), 2.1e7 (ohms)
UP Select next higher measurement range
DOWN Select next lower measurement range
2425 and 2430 DC Mode
<n> = -3.15 to 3.15 Expected reading in amps
-105 to 105 Expected reading in volts
0 to 2.1e7 Expected reading in ohms
DEFault 1.05e-4 (amps), 21 (volts), 2.1e5
(ohms)
MINimum -3.15 (amps), -105 (volts), 0 (ohms)
MAXimum 3.15 (amps), 105 (volts), 2.1e7 (ohms)
UP Select next higher measurement range
DOWN Select next lower measurement range