2400 Series SourceMeter
®
User’s Manual Common Commands 16-5
*SAV, *RCL programming example
Table 16-3 summarizes the basic command sequence for saving and recalling a setup. The
present setup is stored in memory location 2, GPIB defaults are restored, and the memory
location 2 setup is recalled.
*RST — reset
When the *RST command is sent, the SourceMeter performs the following operations:
• Returns the SourceMeter to the GPIB default conditions (see “Default” column of
SCPI tables in Section 18).
• Cancels all pending commands.
• Cancels response to any previously received *OPC and *OPC? commands.
*TRG — trigger
Use the *TRG command to issue a GPIB trigger to the SourceMeter. It has the same effect
as a group execute trigger (GET).
Use the *TRG command as an event to control operation. The SourceMeter reacts to this
trigger if BUS is the programmed arm control source. The control source is programmed
from the TRIGger subsystem.
NOTE Details on triggering are covered in Section 11.
Table 16-3
*SAV, *RCL programming example
Command Description
*SAV 2
*RST
*RCL 2
Save present setup in memory location 2.
Restore GPIB defaults.
Recall location 2 setup.
Return SourceMeter to GPIB defaults
Send bus trigger to SourceMeter
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com