EasyManuals Logo

Keithley SourceMeter 2410 User Manual

Keithley SourceMeter 2410
594 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #571 background image
F-18 Contact Check Function 2400 Series SourceMeter
®
User’s Manual
FAIL?
:CALCulate2:LIMit4:FAIL? Read LIMIT 4 test result
Description This command is used to read the results of the LIMIT 4 test:
0 = Limit 4 test passed
1 = Limit 4 test failed
Reading the results of the limit test does not clear the fail indication of
the test. A failure can be cleared by sending a :SYSTem:CLEar
command.
SOURce2 <NRf> |<NDN>
:CALCulate2:LIMit4:SOURce2 <NRf> |<NDN> Specify pattern; LIMIT 4 failure
Parameters <NRf> = 0 to 7 (3-bit) Decimal value
0 to 15 (4-bit) Decimal value
<NDN> = 0 to #b111 (3-bit) Binary value
0 to #b1111 (4-bit) Binary value
0 to #q7 (3-bit) Octal value
0 to #q17 (4-bit) Octal value
0 to #h7 (3-bit) Hexadecimal value
0 to #hF (4-bit) Hexadecimal value
Query :SOURce2? Query source value for specified limit
Description This command is used to define the digital output fail patterns for the
Limit 4 test (0 to 7, 3-bit; 0 to 15, 4-bit).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2410

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2410 and is the answer not in the manual?

Keithley SourceMeter 2410 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2410
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals