11-18 Triggering 2400 Series SourceMeter
®
User’s Manual
Remote trigger commands
Table 11-1 summarizes remote trigger commands. These commands are covered in more
detail in Section 18 except for *TRG, a common command covered in Section 16.
Table 11-1
Remote trigger commands
Command Description
:INITiate
:ABORt
:ARM:COUNt <n>
:ARM:SOURce <name>
:ARM:TIMer <n>
:ARM:DIRection <name>
:ARM:ILINe <NRf>
:ARM:OLINe <NRf>
:ARM:OUTPut <event list>
:TRIGger:CLEar
:TRIGger:COUNt <n>
:TRIGger:DELay <n>
:TRIGger:SOURce <name>
:TRIGger:DIRection <name>
:TRIGger:ILINe <NRf>
:TRIGger:OLINe <NRf>
:TRIGger:INPut <event list>
:TRIGger:OUTPut <event list>
*TRG
Take SourceMeter out of idle state.
Abort operation, return to idle.
Set arm count (n = count).
Specify arm control source. Name = IMMediate, TLINk, TIMer,
MANual, BUS, NSTest, PSTest, or BSTest.
Set arm layer timer interval (n = interval).
Control arm bypass. (Name = SOURce or ACCeptor).
Select arm layer input line. (NRf = input line #).
Select arm layer output line (NRf = output line #).
Select arm layer output events. (Event list = TENTer, TEXit, or NONE).
Clear any pending input triggers immediately.
Set trigger count (n = count).
Set trigger delay (n = delay).
1
Specify trigger control source. Name = IMMediate or TLINk.
Control trigger bypass. (Name = SOURce or ACCeptor).
Select trigger layer input line (NRf = input line).
Select trigger layer output line (NRf = output line).
Select trigger input layer events. (Event list = SOURce, DELay, SENSe,
or NONE).
Select trigger layer output events. (Event list = SOURce, DELay,
SENSe, or NONE).
Trigger SourceMeter (if BUS source selected).
1. Invalid (ignored) for the Model 2430 Pulse Mode.