EasyManuals Logo
Home>Keithley>Measuring Instruments>SourceMeter 2410

Keithley SourceMeter 2410 User Manual

Keithley SourceMeter 2410
594 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #374 background imageLoading...
Page #374 background image
2400 Series SourceMeter
®
User’s Manual Common Commands 16-7
*WAI — wait-to-continue
Effectively, the *WAI command is a No-Op (no operation) for the SourceMeter and thus,
does not need to be used.
Two types of device commands exist:
Sequential commands — A command whose operations are allowed to finish
before the next command is executed.
•Overlapped commands — A command that allows the execution of subsequent
commands while device operations of the Overlapped command are still in
progress.
The *WAI command is used to suspend the execution of subsequent commands until the
device operations of all previous Overlapped commands are finished. The *WAI command
is not needed for Sequential commands.
Wait until previous commands are completed