User’s Manual Range, Digits, Speed, and Filters 7-13
Filter programming example
Table 7-5 summarizes the command sequence to program filter aspects as follows:
Filter type: moving
Filter count: 20
Filter state: on
Table 7-5
Filter programming example
Command Description
*RST
:SENS:AVER:TCON MOV
:SENS:AVER:COUN 20
:SENS:AVER ON
Restore RST defaults.
Moving average type filter.
Filter count = 20.
Enable filter.