User’s Manual SCPI Command Reference 18-99
Soak time
SOAK <NRf>
:SOURce[1]:SOAK <NRf> Set multiple mode soak time
Parameters <NRf> = soak time (s) 0.000 to 9999.999s
Query :SOAK? Query multiple mode soak time
Description With SYST:RCMode set to MULTiple, SOUR:SOAK specifies the
amount of time after the first point of a sweep that the unit will sit in a
loop actively autoranging up and down to allow a multiple SourceMeter
configuration to settle. (See “SYSTem subsystem,” page 18-106.) This
process will be done only once per INIT, READ?, or MEAS? command.
The soak time is especially useful for low current measurements when
multiple down-range changes from the higher ranges are required.
NOTE For the Model 2430 Pulse Mode, soak time is not used for sweeps.
Pulse Mode delays (Model 2430 only)
WIDth <n>
:SOURce[1]:PULSe:WIDTh <n> Specify pulse width
Parameters <n> = 0.00015 to 0.00500 Specify pulse width in seconds
MINimum 0.00015 seconds
MAXimum 0.00500 seconds
DEFault 0.00015 seconds
Query :WIDTh? Query pulse width
:WIDTh? DEFault Query *RST default pulse width
:WIDTh? MINimum Query lowest allowable pulse width
:WIDTh? MAXimum Query highest allowable pulse width
Description This command is used to set the pulse width for the Model 2430 Pulse
Mode. The pulse width is the time that the output remains on, including
the signal measurement.
The valid range of pulse width periods is from 0.15ms to 5ms. The pulse
width can be set to <0.15ms, however, 0.15ms is the minimum pulse
width that will occur. On the 10A range (source or measure), the pulse
width can be set up to 5ms, however, it will be limited to 2.5ms.
NOTE See Section 5 for details on Model 2430 Pulse Mode.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2410

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2410 and is the answer not in the manual?

Keithley SourceMeter 2410 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2410
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals