2400 Series SourceMeter
®
User’s Manual SCPI Command Reference 18-51
Data elements
ELEMents <item list>
:FORMat:ELEMents [SENSe[1]] <item list> Specify data elements for data string
Parameters <item list> = VOLTage Includes voltage reading
CURRent Includes current reading
RESistance Includes resistance reading
TIME Includes timestamp
STATus Includes status information
NOTE Each item in the list must be separated by a comma (i.e., :ELEMents, VOLTage,
CURRent, RESistance).
Query :ELEMents? Query elements in data string
This command is used to specify the elements to be included in the data
string in response to the following queries:
:FETCh?
:READ?
:MEASure?
:TRACe:DATA?
You can specify from one to all five elements. Each element in the list
must be separated by a comma (,). These elements (shown in Figure
18-1) are explained as follows:
NOTE An overflow reading reads as +9.9E37.
VOLTage — This element provides the voltage measurement or the pro-
grammed voltage source reading. If sourcing voltage and measuring
voltage, this element will provide the voltage measurement (measure
reading takes priority over source reading). If voltage is not sourced or
measured, the NAN (not a number) value of +9.91e37 is used.
CURRent — This element provides the current measurement or the
programmed current source reading. If sourcing current and measuring
current, this element will provide the current measurement (measure
reading takes priority over source reading). If current is not sourced or
measured, the NAN (not a number) value of +9.91e37 is used.
RESistance — This element provides the resistance measurement. If
resistance is not measured, the NAN (not a number) value of +9.91e37
is used.
TIMEA timestamp is available to reference each group of readings
to a point in time. The relative timestamp operates as a timer that starts
at zero seconds when the instrument is turned on or when the relative
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2410

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2410 and is the answer not in the manual?

Keithley SourceMeter 2410 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2410
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals