18-94 SCPI Command Reference 2400 Series SourceMeter
®
User’s Manual
of the SourceMeter with CALC1 disabled. User-defined math expressions are
replaced with the Power math expression.
Error 809 (Source memory location revised) occurs when a memory sweep ref-
erences an expression that no longer exists. The memory sweep is revised to dis-
able CALC1.
In order to execute a memory sweep, the product of the arm count and trigger
count should be at least the same as the number of points in the sweep.
Use the :FUNCtion:MODE command to select the MEMory function. (See “Select func-
tion modes,” page 18-73.) Arm count (ARM:COUNt) and trigger count
(TRIGger:COUNt) are set from the Trigger Subsystem.
SAVE <NRf>
:SOURce[1]:MEMory:SAVE <NRf> Save setup in specified memory location
Parameters <NRf> = 1 to 100 Specify memory location
Description This command is used to save the present instrument setup in a memory
location. Up to 100 setups can be saved. The following settings are
saved in each source memory location:
SENSe[1]:CURRent:NPLCycles
SENSe[1]:RESistance:NPLCycles
SENSe[1]:VOLTage:NPLCycles
SENSe[1]:FUNCtion:CONCurrent
SENSe[1]:FUNCtion:ON
SENSe[1]:FUNCtion:OFF
SENSe[1]:RESistance:MODE
SENSe[1]:RESistance:OCOMpensated
SENSe[1]:AVERage:STATe
SENSe[1]:AVERage:TCONtrol
SENSe[1]:AVERage:COUNt
SOURce[1]:FUNCtion:SHAPe (Model 2430 only)
SOURce[1]:FUNCtion:MODE
SOURce[1]:DELay (Model 2430 in DC Mode)
SOURce[1]:DELay:AUTO (Model 2430 in DC Mode)
SOURce[1]...X...:TRIGgered:SFACtor
SOURce[1]...X...:TRIGgered:SFACtor:STATe
where: ...X... = :CURRent or :VOLTage (based on source mode)
SOURce[1]:PULSe:WIDTh (Model 2430 Pulse Mode only)
SOURce[1]:PULSe:DELay (Model 2430 Pulse Mode only)
Source Value, Range, Auto Range
Sense Protection, Range, Auto Range
SYSTem:AZERo:STATe
SYSTem:RSENse