7-4 Adjustments
and white strips in the middle or around are clear, while the adjacent strips
can be distinguished to pass the test;
3.
AshRank
Images of different gray rank levels can be distinguished easily with a
smooth transition, and the brightness transition can also be obtained from
the images. Gray rank of low level is not obviously lean to red or green, and
then the test is passed.
4.
ColorRank
Images of different color rank levels can be distinguished easily with a
smooth transition, and the brightness transition can also be obtained from
the images.
5.
Colorful
If different kinds of color graphics appear on the screen where the color and
the word of each graphics are consistent, test is passed.
6.
focus
Press [Set] to switch among the 3 interfaces, if the images are clear with
letters and characters easy to be recognized and there is no ambiguous
display or variance among them, the test is passed;
7.
BadPoint
Press [Set] to enter pure color interfaces of green, blue, red, black and white.
Observe the LCD screen. Check bright point and dark point in the black and
white interfaces. Criterion for pass: bright point number is 0; number of
continuous dark point pair is ≤3, and no continuous dark point appears in the
image area; there are no three or more than three continuous dark point;
dark point defects number is ≤7, and number of dark points in the image area
is ≤2; point flash defects is 0; space between fail points is ≥5mm.
4. Click [Exit] or press [ESC] to return to the main screen interface.
7.1.4 Monitor Parameter Setting
NOTE:
After changing the main monitor, parameter loading should be performed to match the
monitor and the system.
The parameters of the main monitor include color temp, brightness/contrast, Gamma, and
Default, information of loading Default parameters are introduced here:
1. Press <1> and <4> key on the monitor as shown in the figure for more than 3 seconds to enter
the screen.

Table of Contents

Other manuals for Mindray DC-8 EXP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-8 EXP and is the answer not in the manual?

Mindray DC-8 EXP Specifications

General IconGeneral
BrandMindray
ModelDC-8 EXP
CategoryMedical Equipment
LanguageEnglish

Related product manuals