Software Installation &Maintenance 6-3
6.4 Software maintenance
6.4.1 Product Configuration
NOTE:
[Config] is available on the Maintenance menu only if the operator logged on the
system as Service.
1. Make PCF file according to the product (made in the factory, for special models, the PCF file is
already archived), then copy it to the USB disk. Insert the USB disk into the USB port; click
[Config] on the maintenance menu to open the following dialogue box.
2. Select correct PCF files, and click [OK]. After the operation is completed, the successful prompt
appears. Restart the system and the configuration takes effect.
Special configurations including:
Dedicated product LOGO, hospital LOGO, dormant/screen saver logo, image active/inactive
symbols in dual B mode, system information, system start-up/shut down screen, etc.
NOTE:
1.
System general configuration is already configured during the resume
package restoration. No manual configuration is required if no special
requirement.
2.
Don’t change the model of the machine; otherwise the key files may be lost.
6.4.2 Log Maintenance
6.4.2.1 Export the Log
NOTE:
This operation can be performed by both internal and external users.

Table of Contents

Other manuals for Mindray DC-8 EXP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-8 EXP and is the answer not in the manual?

Mindray DC-8 EXP Specifications

General IconGeneral
BrandMindray
ModelDC-8 EXP
CategoryMedical Equipment
LanguageEnglish

Related product manuals