Electrical Safety Inspection A-1
Appendix A Electrical Safety
Inspection
The following electrical safety tests are recommended as part of a comprehensive preventive
maintenance program. They are a proven means of detecting abnormalities that, if undetected,
could prove dangerous to either the patient or the operator. Additional tests may be required
according to local regulations.
All tests can be performed using commercially available safety analyzer test equipment. These
procedures assume the use of a 601PRO
XL
International Safety Analyzer or equivalent safety
analyzer. Other popular testers complying with IEC 60601-1 used in Europe such as Fluke, Metron,
or Gerb may require modifications to the procedure. Follow the instructions of the analyzer
manufacturer.
The consistent use of a safety analyzer as a routine step in closing a repair or upgrade is
emphasized as a mandatory step if an approved agency status is to be maintained. The safety
analyzer also proves to be an excellent troubleshooting tool to detect abnormalities of line voltage
and grounding, as well as total current loads.

Table of Contents

Other manuals for Mindray DC-8 EXP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-8 EXP and is the answer not in the manual?

Mindray DC-8 EXP Specifications

General IconGeneral
BrandMindray
ModelDC-8 EXP
CategoryMedical Equipment
LanguageEnglish

Related product manuals