12-6 Care & Maintenance
5. Clean Probe
Tools: mild soapy water , dry soft cloth, soft brush
Method:
a) Wipe out the dust attached to surface of probe head, connector and cable.
b) Use soft brush to brush the dust inside probe connector gently.
c) Remained stain or dust attached to surface of cable or surface of connector should be
washed out by cloth with a little soapy water, and then air-dry.
Note: Don’t use cloth with water to clean the probe connector.
6. Clean Holders
Tool: dry soft cloth , soapy water, soft brush
Method:
Surface of power button should be cleaned by dry soft cloth. Remained stain should be washed
out by cloth with clear water or soapy water (Note: cloth should not be dipped with too much
water, which may cause electric shock), and then air-dry the surface.
a)
Use dry soft cloth to wipe off the dust attached to inside, outside or gap of probe holder or
gel holder. As to small intra-cavity probe holder or its gap, use the soft brush to brush the
dust or stain.
b) Remained stain attached to inside, outside of holder should be washed out by cloth with a
little soapy water and then air-dry after take it out.
c) Gel heater: Take the gel heater out after pull out the power supply cable, use dry soft cloth
to wipe off the dust attached to inside and outside, then brush the dust in the ostiole of gel
heater or brush the stain with a little soap water and air-dry at last.
7. Clean Cover
Tools: dry soft cloth, soapy water
Method:
Use dry soft cloth to clean the cover of the system. If the system is dirty, moisten the soft
cloth with mild soapy water and wipe off any stains, then air-dry.
Note: Be sure to use soft brush to brush the dust attached to all the sockets or interfaces which
can be seen (such as probe sockets, sockets or interfaces in IO panel and power supply
panel),not the cloth with water.
12.2.2 Clean the Peripherals
Do the cleaning maintenance according to your actual peripheral configuration; items which are not
configured can be skipped.