9-24 Structure and Assembly/Disassembly
9.3.10 Monitor Assembly
NOTE:
Parameters must be reloaded after changing monitor assembly. Monitor can be normally
used only after the matching of machine and monitor. Refer to 7.1.4 for details.
1. Adjust the monitor to the upright position and fix it. Unscrew two M4X12 screws, remove the
rear cover.
2. Unscrew three M4X8 screws to remove signal cable of monitor and power wire.
3. Adjust the display to the upright position and fix it. Unscrew six M4X12 screws to remove
monitor assembly.