12-8 Care & Maintenance
No.
Content
Method
4.
Battery
Check the battery periodically
a) Check if battery can be charged normally when power-on: That
the current capacity is 100% or capacity increases after a short
time indicates that the battery can be charged normally. It takes
less than 5 minutes to increase 1% capacity when the total
capacity is less than 90% and it takes more time when the
capacity is more than 90%.
b) Disconnect the system from the AC power supply to confirm if
the system can maintain normal standby status in the battery
power supply by inspecting standby status indicator, That the
standby status indicator blinks in orange continually indicates
that standby status is normal.
12.3.2 System Function Check
The system function checking is not required during Preventive Maintenance. Engineer or
Customer may use it as part of their product Quality Assurance Program tests
Table 12-6 System function list
No.
Content
Method
1.
B mode
Verify basic operation of B mode. Check basic software and
hardware controls affecting B mode operations.
2.
Color mode
Verify basic operation of Color mode. Check basic software
and hardware controls affecting Color mode operations.
3.
Doppler mode
(PW/CW)
Verify basic operation of Doppler mode. Check basic software
and hardware controls affecting Doppler mode operations.
4.
M mode
Verify basic operation of M mode. Check basic software and
hardware controls affecting M mode operations.
5.
Measurement2D, M,
Doppler general
measurement, applied
measurement
optional
Scanning gray scale imaging on phantom, verify distance and
area accuracy with measurement control. Verify measurement
accuracy by performance test.
6.
Keyboard test
Operate keyboard test to verify if all control keys can work
normally.
7.
LCD
Verify LCD display function and parameters adjustment. Refer
to that of LCD checking.
8.
Software menu check
Verify software menu display function: if each operation menu
and page can be accessed.
RemarkPlease refer to 5.45.5 for details.

Table of Contents

Other manuals for Mindray DC-8 EXP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-8 EXP and is the answer not in the manual?

Mindray DC-8 EXP Specifications

General IconGeneral
BrandMindray
ModelDC-8 EXP
CategoryMedical Equipment
LanguageEnglish

Related product manuals