C-2 Description of Self-diagnosis Test Items
The interface can’t be closed manually after the test is started, until the test is completed. If the test
failed, please turn off the power as soon as possible.
Suggestion when test failed:
Check if PC module heat sink was correctly installed, if not, reinstalled it.
C.1.3 Z0301 PC Module and MF FPGA Interconnection Test
Test Content
Test if the LPC communication between PC module and MF FPGA is normal.
Suggestion when test failed:
1. Check if the following listed hardware and the related circuits are normal, check if the interfaces
are well welded and if the module is working normally.
2. Replacing the digital board is recommended.
C.1.4 Z0302 SVIDEO Buffer DDRII Test
Test Content
The super stratum test item: Z0201 PC Module and MF FPGA Interconnection Test
Test if SVIDEO DDR II works normally.
Suggestion when test failed:
1. Check if the following listed hardware and the related circuits are normal, check if the interfaces
are well welded and if the module is working normally.
2. Replacing the digital board is recommended.