Emerson Process Management GmbH & Co. OHG5-24
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
5.7 Checking the Settings
You may turn the alarm function On or Off
separately for each channel ("Alarm Monitor").
It´s also possible to use some of the thres-
holds only. In this case, set the not used to
a level outside the range limits (for this, "Lo"
and "LoLo" support entering negative values).
In case of an alarm, you may like to enter
the STATES submenu, to check which one
is triggered.
The most important settings have now been
checked and the unit is congured to suite
your needs.
A backup copy of these conguration data
can now be saved.
Press
left until SETUP and then open SAVE-
LOAD.
5.7.6 Backup the Settings
This menu gives you the choice, to either
make a local backup to a protected memory
area
restore the factory default settings, or
make a backup to an external USB device.
Ch1
Concentration
Alarm Monitor: On
LoLo Level: 50.000 ppm
Lo Level: 100.000 ppm
Hi Level: 400.000 ppm
HiHi Level: 600.000 ppm
Hysteresis: 10.000 ppm
States..
Save-Load
Local Backup..
Factory Defaults..
USB Backup..
USB Firmware Update..

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson X-STREAM XE and is the answer not in the manual?

Emerson X-STREAM XE Specifications

General IconGeneral
BrandEmerson
ModelX-STREAM XE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals