7.6.4 Replacing the Trace Oxygen Sensor
Replacing the trace oxygen sensor requires
considering special instructions, shipped to-
gether with every single sensor.
Two versions of this sensor are available,
differing in the background gas they can be
used for:
P/N 427.9102: For sample gases containing
acid gases, hydrocarbons or hydrogen
P/N 427.9103: For sample gases without acid
gases, hydrocarbons and hydrogen
If replacing this sensor is necessary (con-
sumable), contact Emerson for information
on how to order a substitute.
A new sensor will be provided together with
detailled installation and handling instructions.
Consider all information given
by the replacement instructions
to avoid damaging the sensor,
and to achieve best possible
life time!
7.6 Replacing Worn Out Sensors