Maintenance & Procedures
7.4.3 Advanced Calibration
Fig. 7-5: Span All Calibration Procedure Flow Diagram
start to
span all chan-
nels with the cur-
rent span gas valve
assigned, conside-
ring the individual
purge times
select
channel 1
increase
channel no.
by 1
wait until
all selected
channels are
spanned
select all
channels with
the same
span gas valve
assigned
is current
channel the
last channel?
is this
channel
already
spanned?
start
span all
procedure
END
Yes
Yes
No
No