7.11 Web Browser
IP address for Ethernet2 connector
To directly gain access to the instrument´s web
browser interface, rst ensure the instrument
is powered. Next connect it to the computer's
network adapter via Ethernet2 connector (see
Fig. 7-22) by means of an Ethernet cross-
over cable. Standard Ethernet cables do
NOT support direct connections!
For Ethernet2 connector set "Use DHCP"
to No (see gure to the left).
Depending on settings possibly carried out
earlier, the analyzer may now show an IP (see
gure to the left). If so, setup your computer's
IP the same way, only differing in the last
group (here e.g. to 123.456.78.10).
Alternatively you may use the analyzer's xed
IP, that is not shown on any menu page and
is accessible via Ethernet2 only:
The analyzer is now assigned the IP
192.168.1.88.
Congure your computer's IP to meet the
same net (192.168.1.) and assign a new
IP (e. g. 192.168.1.10). Do not use the
same IP as is assigned to the analyzer.
7.11.2 Connection to Single Computer
Conguring an IP Address for Computers Running Microsoft Windows
To congure your computer you need an administrator account!
Go to Start > Control Panel > Network Connections
Right-click on your LAN connection and click "Properties"
Under the heading "This connection uses the following items", click "Internet Protocol (TCP/IP)"
Click "Properties"
A new window should pop up, click "Alternate Conguration"
Click "User Congured" radio button
Setup the IP Address as 192.168.1.10, Subnet Mask as 255.255.255.0 and Default Gateway as 192.168.1.1.
Click "OK"
Click "Close" in "LAN connection properties".
On your computer open a web browser and
enter the instrument´s IP address. If everything
is congured properly, the analyzer's logon
screen shows up ( next page ).
Ethernet2
MAC 1234567.10
IP: 123.456.78.9
Subnet: 255.255.255.1
Gateway: 123.456.78.0
Use DHCP: No
IP Status Ready
Apply Conguration!
Setup..
Communcation..
Ethernet2..

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson X-STREAM XE and is the answer not in the manual?

Emerson X-STREAM XE Specifications

General IconGeneral
BrandEmerson
ModelX-STREAM XE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals