| CARDIOHELP System | 5 Service Interfaces | 87 |
Service Manual | 3.3 | EN | 05
Copyright Maquet Cardiopulmonary GmbH
u Release Version __________.
13 Carry out software validation on the device (⇨ "Validate the Software", page 80).
14 Restore battery change date and customer settings according to the User‘s Manual (⇨ "Saving the
Battery Change", page 92).
15 Carry out a complete inspection (⇨ "Inspection", page 33).
u Software update completed.
u Inspection has been carried out.
5.3.3 Analyzing Events and Errors
For the analysis of events and faults, the following options are available to the service technician.
[Service pool] symbol
n Reading out, analyzing saving and deleting events and errors of the service pool.
[User pool] symbol
n Reading out, analyzing saving and deleting events and errors of the user pool.
[Open event] symbol
n Open and evaluate old events lists
1 “Status” window
2 “Event List” window
3 Events display window
4 “Reading” selection area
5 “Analysis” selection area
6 “Read now” button
7 “CLEAR!” button
8 “OK” button
Service events screen
Read Out Events
1 Read out the data by clicking on the appropriate button [Service-
Pool] or [User-Pool].
Read out events
2 Select the option [Read all events] or [Read only] with the desired
amount of data (e.g., 500).
Selection area "Read"
3 Click on the [Read now] button.
[Read now] button
The data is read out.