| 90 | 5 Service Interfaces | CARDIOHELP System |
Service Manual | 3.3 | EN | 05
Copyright Maquet Cardiopulmonary GmbH
5 Click on the [CLEAR!] button.
[Clear] button
The message “Really delete ALL events?” is displayed.
Warning message
6 Click on the [Yes] button.
u Service pool deleted.
u User pool deleted.
5.3.4 Licensing
WARNING!
Carry out final tasks.
To finish the task, the following step must be carried out in order to ensure error-free use.
n Carrying out the inspection (⇨ "Inspection", page 33).
1 Activate the service interface (⇨ "Activate the Service Interface", page 79).
2 Start the service tool software (⇨ "Start Service Tool Software", page 85).
3 Switch to the “Special” tab.
[Licensing] symbol
4 Click on the [Licensing] button.
The “Save license” screen is displayed.
Open license
5 Click on the [Open] button.
6 Open the valid license key.
The license key for the CARDIOHELP has the file extension “lic”.
7 Click on the [Send] button.
Sending the license
The license is saved.
The “Licensing successful” screen is displayed.
8 Click on the [OK] button.
9 Click on the [Close] button.
The “Save license” screen is closed.

Table of Contents

Other manuals for Getinge MAQUET CARDIOHELP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge MAQUET CARDIOHELP and is the answer not in the manual?

Getinge MAQUET CARDIOHELP Specifications

General IconGeneral
BrandGetinge
ModelMAQUET CARDIOHELP
CategoryMedical Equipment
LanguageEnglish

Related product manuals