| 94 | 5 Service Interfaces | CARDIOHELP System |
Service Manual | 3.3 | EN | 05
Copyright Maquet Cardiopulmonary GmbH
2 Start the service tool software (⇨ "Start Service Tool Software", page 85).
3 Switch to the “Special” tab.
4 Click on the [Settings] button.
[Settings] symbol
The “Delete Configuration” warning is displayed.
"Delete Configuration"
5 Click on the [Yes] button if you wish to continue.
If you do not want to delete the clinic configuration, click on the [No] button.
The warning “Clinical Configuration deleted” is displayed.
"Clinical Configuration deleted" warning
6 Click on the [OK] button.
5.3.9 Country Settings
WARNING!
The country settings are mandatory for USA.
Activate country settings
1 Activate the service interface (⇨ "Activate the Service Interface", page 79).
2 Start the service tool software (⇨ "Start Service Tool Software", page 85).
3 Switch to the “Special” tab.
4 Click on the [Country settings] button.
[Country settings] symbol
The “Country Settings” pop-up screen is displayed.
Country settings
5 Select the necessary check box.
Mark the [United states of America] check box if the device will be used in USA.
Mark the [Rest of the world] check box for all other countries.
6 Click on the [Save] button.
The message “Country setting activated!“ is displayed.
Country settings activated