| CARDIOHELP System | 5 Service Interfaces | 93 |
Service Manual | 3.3 | EN | 05
Copyright Maquet Cardiopulmonary GmbH
5 Choose the battery change date.
▶ Normally the current date.
▶ Except the battery change date must be restored.
6 Click on the [Save] button.
"Saving the battery change"
▶ The message ”Set Battery Change Date“ is displayed.
"Set Batterie Change Date"
u The battery change is saved in the CARDIOHELP.
5.3.7 Saving the Serial Number
CAUTION!
After replacing the HMI PCB, the serial number of the device must be saved on the new HMI PCB.
1 Activate the service interface (⇨ "Activate the Service Interface", page 79).
2 Start the service tool software (⇨ "Start Service Tool Software", page 85).
3 Switch to the “Special” tab.
4 Click on the [Save Serial No] button.
[Save Serial No] symbol
▶ The “Set Serial Number” screen is displayed.
"Set Serial Number" screen
5 Enter the device serial number of the rating plate.
6 Click on the [Save] button.
▶ The serial number is saved on the HMI PCB.
7 Click on the [Close] button.
u Device serial number is saved on the HMI PCB.
5.3.8 Delete Clinic Configuration - Save Factory Settings
1 Activate the service interface (⇨ "Activate the Service Interface", page 79).