Installation - Closing the MSC Lid
Part No.: 4418309_Rev09 Fusion4 MSC-L
Honeywell Installation & Operation Manual 4 - 11
4.4 Closing the MSC Lid
To close the MSC-L Lid
1. Make sure the O-ring is in place and it is not damaged. If the O-ring
is damaged, replace it first.
2. Remove the 2x cable (grounding cable and cable to prevent over-
stretching door after oppening the door) before replacing the O-ring.
NOTE: The O-ring must be replaced only by the Honeywell
Enraf trained by service technician.
3. Make sure the wires are not caught between the lid and the box
flange.
4. Make sure the ground bonding provided is effective.
5. Tighten all the 32 bolts crosswise using an 8 mm Allen key and the
applicable torque value (33 Nm).
6. Replace the cosmetic cover.