EasyManuals Logo
Home>Keithley>Multimeter>2701

Keithley 2701 User Manual

Keithley 2701
506 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #183 background imageLoading...
Page #183 background image
Model 2701 User’s Manual Range, Digits, Rate, Bandwidth, and Filter 4-17
Figure 4-3
Filter window
+1% of range
-1% of range
A
B
Voltage
t
1 t2 t3 t4 t5 t6 t7 t8 t9 t10 t11
+1% of range
-1% of range
Windows
Violation
Integration
Time
A
1
A1
A1
A1
A1
Rdg
#1
A2
A1
A1
A1
A1
Rdg
#2
A3
A2
A1
A1
A1
Rdg
#3
A4
A3
A2
A1
A1
Rdg
#4
A5
A4
A3
A2
A1
Rdg
#5
A6
A5
A4
A3
A2
Rdg
#6
B1
A5
A4
A3
A2
Rdg
#7
B2
B1
A5
A4
A3
Rdg
#8
B3
B2
B1
A5
A4
Rdg
#9
B4
B3
B2
B1
A5
Rdg
#10
B5
B4
B3
B2
B1
Rdg
#11
Filter configuration:
Type = Moving
Count = 5
Window = None
Conversions:
A1
A1
A1
A1
A1
Rdg
#1
A2
A1
A1
A1
A1
Rdg
#2
A3
A2
A1
A1
A1
Rdg
#3
A4
A3
A2
A1
A1
Rdg
#4
A5
A4
A3
A2
A1
Rdg
#5
A6
A5
A4
A3
A2
Rdg
#6
B1
B1
B1
B1
B1
Rdg
#7
B2
B1
B1
B1
B1
Rdg
#8
B3
B2
B1
B1
B1
Rdg
#9
B4
B3
B2
B1
B1
Rdg
#10
B5
B4
B3
B2
B1
Rdg
#11
Filter configuration:
Type = Moving
Count = 5
Window = 1%
A1
A1
A1
A1
A1
A2
A1
A1
A1
A1
A3
A2
A1
A1
A1
A4
A3
A2
A1
A1
A5
A4
A3
A2
A1
Rdg
#1
A6
A6
A6
A6
A6
B1
A6
A6
A6
A6
B2
B1
A6
A6
A6
B3
B2
B1
A6
A6
B4
B3
B2
B1
A6
Rdg
#2
B5
B5
B5
B5
B5
Filter configuration:
Type = Repeating
Count = 5
Window = None
A1
A1
A1
A1
A1
A2
A1
A1
A1
A1
A3
A2
A1
A1
A1
A4
A3
A2
A1
A1
A5
A4
A3
A2
A1
Rdg
#1
A6
A6
A6
A6
A6
B1
B1
B1
B1
B1
B2
B1
B1
B1
B1
B3
B2
B1
B1
B1
B4
B3
B2
B1
B1
Rdg
#2
B5
B4
B3
B2
B1
Filter configuration:
Type = Repeating
Count = 5
Window = 1%

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
BrandKeithley
Model2701
CategoryMultimeter
LanguageEnglish

Related product manuals