2-14 Close/Open Switching Module Channels Model 2701 User’s Manual
Remote programming example (system channel operation)
The following example assumes a Model 7700 installed in slot 1, and the 4 function of
the Model 2701 is selected. This command sequence connects channel 101 and its paired
channel (111) to DMM Input and Sense as shown in Figure 2-2.
ROUT:OPEN:ALL ' Open all channels.
ROUT:CLOS (@101) ' Close channels 101, 111, 123, 124, and 125.
Non-amp and non-measure switching modules
There are Keithley switching modules that do not support current measurements and there
are modules that do not support any measurements at all.
Non-amps moduleWith an amps function selected (DCI or ACI), system channel
operation cannot be used to close channels on that module.
Non-measure module — For front panel operation, system channel operation cannot be
used to close channels. For remote programming, system channel operation can be used,
but only the one specified channel will close. All other channels on the module will open.
Non-amps switching modules
NOTE Presently, non-amps Keithley modules include the Models 7701, 7703, 7706,
7707, 7708, and 7709. You can check the Keithley website (www.keithley.com)
for new modules.
A non-amp module does not support amps measurements. System channel operation can-
not be used to close channels while an amps function (DCI or ICI) is selected.
If an amps function (DCI or ACI) is selected and you attempt to close a system channel,
the message “NO AMPS CHAN” will be displayed briefly. For remote programming,
error -222 (Parameter data out of range) is generated. Example:
SYST:PRES ' Restores system preset defaults.
SENS:FUNC CURR:DC ' Selects DCI function.
ROUT:CLOS (@101) ' Attempts to close system channel 101 Generates
' error -222.
If a system channel is already closed and you attempt to select the DCI or ACI function,
the message “INVALID FUNC” will be displayed briefly. For remote programming,
error -221 (Settings conflict) is generated. Example:
SYST:PRES ' Restores system preset defaults.
ROUT:CLOS (@101) ' Close system channel 101.
SENS:FUNC CURR:DC ' Attempts to select DCI function Generates
' error -221.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
BrandKeithley
Model2701
CategoryMultimeter
LanguageEnglish

Related product manuals