Model 2701 User’s Manual Status Structure 11-23
On power-up, all error messages are enabled and will go into the Error Queue as they
occur. Status messages are not enabled and will not go into the queue. As listed in
Table 11-7, there are commands to enable and/or disable messages. For these commands,
the <list> parameter is used to specify which messages to enable or disable. The messages
are specified by their codes. The following examples show various forms for using the
<list> parameter.
<list> = (-110) Single message
= (-110:-222) Range of messages (-110 through -222)
= (-110:-222, -220) Range entry and single entry (separated by a comma)
When you enable messages, messages not specified in the list are disabled. When you
disable messages, each listed message is removed from the enabled list.
NOTE To prevent all messages from entering the Error Queue, send the enable
command along with the null list parameter as follows:
STATus:QUEue:ENABle ().
Programming example read error queue
The following example can be run from the KE2700 Instrument Driver using the example
named “ReadErrorQueue” in Table G-1 of Appendix G.
The following command reads the error queue:
STAT:QUE?
Table 11-7
Error queue commands
Command Description Notes
STATus:QUEue:[:NEXT]? Read and clear oldest error/status (code and message). 1
STATus:QUEue:ENABle <list> Specify error and status messages for Error Queue. 2
STATus:QUEue:ENABle? Read the enabled messages.
STATus:QUEue:DISable <list> Specify messages not to be placed in queue. 2
STATus:QUEue:DISable? Read the disabled messages.
STATus:QUEue:CLEar Clear messages from Error Queue.
SYSTem:ERRor? Read Error Queue. 1
SYSTem:CLEar Clear messages from Error Queue.
Notes:
1. Power-up and *CLS empties the Error Queue. STATus:PRESet has no effect.
2. Power-up enables error messages and disables status messages. *CLS and STATus:PRESet have no effect.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
BrandKeithley
Model2701
CategoryMultimeter
LanguageEnglish

Related product manuals