Model 2701 User’s Manual Signal Processing Sequence and Data Flow D-7
Data ow (remote operation)
Remote operation can be used with triggering configured to perform a specified number of
measurements and then stop. The various read commands (SENS:DATA?, FETCh?,
READ?, MEAS?, CALC2:DATA?, TRACe:DATA?, and CALC1:DATA?) return the data
array(s) acquired during the measurement cycle. Data flow for this triggering
configuration is summarized by the block diagram shown in Figure D-4. Refer to this
block diagram for the following discussion.
Figure D-4
Data flow for remote operation
SENSe
Measurement, Filter,
Rel, and Ratio
or Ch Avg
CALC1
CALC3
Limit Tests
TRACe:DATA?
CALC2
Min, Max, Sdev
Mean, Pk-Pk
[SENS[1]]:DATA[:LAT]?
[SENS[1]]:DATA:FRESh?
Math (mX+B,
Percent,
or 1/X)
CALC3:LIM1:FAIL?
CALC3:LIM2:FAIL?
FETCh?
READ?
MEAS?
CALC[1]:DATA[:LAT]?
CALC[1]:DATA:FRESh?
Sample
Buffer
TRACe
Data Store
TRAC:CLE
INIT:CONT OFF
TRIG:COUN 1
INIT
CALC2:IMM?
CALC2:IMM
CALC2:DATA?

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
BrandKeithley
Model2701
CategoryMultimeter
LanguageEnglish

Related product manuals