Model 2701 User’s Manual Range, Digits, Rate, Bandwidth, and Filter 4-11
Remote programming — rate and bandwidth
Rate and bandwidth commands
The commands to set the integration rate and bandwidth are listed in Table 4-5. Additional
information on these commands follows the table.
NOTE Query commands are not included in Table 4-5. All commands for the SENSe
subsystem are provided in Table 15-5.
Table 4-5
Rate and bandwidth commands
Commands
1, 6
Description Default
Integration rate commands
[SENSe[1]]
:VOLTage[:DC]:NPLCycles <n> [, <clist>] Set rate for DCV in PLCs; <n> = 0.002 to x.
2
5.0
:VOLTage[:DC]:APERture <n>
[, <clist>]
Set rate for DCV in secs; <NRf> = 3.333e-5 to 1. (Note 3)
:VOLTage:AC:NPLCycles <n> [, <clist>] Set rate for ACV in PLCs; <n> = 0.002 to x.
2,4
5.0
:VOLTage:AC:APERture <n> [, <clist>] Set rate for ACV in secs; <n> = 3.333e-5 to 1.
4
(Note 3)
:CURRent[:DC]:NPLCycles <n> [, <clist>] Set rate for DCI in PLCs; <n> = 0.002 to x.
2
5.0
:CURRent[:DC]:APERture <n> [, <clist>] Set rate for DCI in secs; <n> = 3.333e-5 to 1. (Note 3)
:CURRent:AC:NPLCycles <n> [, <clist>] Set rate for ACI in PLCs; <n> = 0.002 to x.
2,4
5.0
:CURRent:AC:APERture <n> [, <clist>] Set rate for ACI in secs; <n> = 3.333e-5 to 1.
4
(Note 3)
:RESistance:NPLCycles <n> [, <clist>] Set rate for Ω2 in PLCs; <n> = 0.002 to x.
2
5.0
:RESistance:APERture <n> [, <clist>] Set rate for Ω2 in secs; <n> = 3.333e-5 to 1. (Note 3)
:FRESistance:NPLCycles <n> [, <clist>] Set rate for Ω4 in PLCs; <n> = 0.002 to x.
2
5.0
:FRESistance:APERture <n> [, <clist>] Set rate for Ω4 in secs; <n> = 3.333e-5 to 1. (Note 3)
:TEMPerature:NPLCycles <n> [, <clist>] Set rate for TEMP in PLCs; <n> = 0.002 to x.
2
5.0
:TEMPerature:APERture <n> [, <clist>] Set rate for TEMP in secs; <n> = 3.333e-5 to 1. (Note 3)