15-8 SCPI Reference Tables Model 2701 User’s Manual
Table 15-5
SENSe command summary
Command Description
Default
parameter Ref SCPI
[SENSe[1]]
:FUNCtion <name>
[, <clist>]
Select function: ‘VOLTage[:DC]’,
‘VOLTage :AC’, ‘CURRent[:DC]’,
‘CURRent:AC’, ‘RESistance’,
‘FRESistance’, ‘TEMPerature’,
‘FREQuency’, ‘PERiod’, ‘CONTinuity’.
VOLT:DC Sec 3 ✓
:FUNCtion? [<clist>] Query function. ✓
:DATA[:LATest]? Return the last reading. Sec 3 ✓
:DATA:FRESh? Return the last “fresh” reading. Sec 3
:HOLD Path to control Hold feature: Sec 8
:WINDow <NRf> Set Hold window in % (0.01 to 20). 1
:WINDow? Query Hold window.
:COUNt <NRf> Set Hold count; 2 to 100. 5
:COUNt? Query Hold count.
:STATe <NRf> Enable or disable Hold. OFF
:STATe? Query state of Hold.
:CAVerage Channel average calculation: Sec 5
:DELay <NRf> [, <clist>] Set delay between the two measurements
in seconds (0 to 99999.999).
1
0.5
:DELay? [<clist>] Query delay.
[:STATe] <b> [, <clist>] Enable or disable channel average. OFF
[:STATe]? [, <clist>] Query state of channel average.
:RATio Channel ratio calculation: OFF Sec 5
:DELay <NRf> [, <clist>] Set delay between the two measurements
in seconds (0 to 99999.999).
1
0.5
:DELay? [<clist>] Query delay.
[:STATe] <b> [, <clist>] Enable or disable channel ratio. OFF
[:STATe]? [, <clist>] Query state of channel ratio.