10-16 Remote Operations Model 2701 User’s Manual
Internal web page
The Model 2701 has an internal web page (see Figure 10-9). The web page is organized as
follows:
• Diagnostic summary:
• Ethernet details – You can view and/or change the Ethernet settings of the
Model 2701.
• Instrument details – Provides read-only information about the Model 2701
(revision levels, serial numbers, and calibration status).
• Card details — Provides read-only information about switching modules
installed in the mainframe (revision level, serial number, and calibration
status).
• Password details — Allows the user to enable (yes) or disable (no) the use of
password. When enabled, the password can be changed.
• Interactive control panel — From the control panel, the user can send commands
and queries to Model 2701. Also, with the click of button, the instrument will
continuously take readings. Readings, responses to queries, and any error messages
are displayed in the control panel.
Opening the web page
System requirements
Internet Explorer (IE 5.0 or greater) is needed to open the 2701 web page. Make sure the
PC and Model 2701 is properly set for Ethernet operation. Details on using the Ethernet is
provided in the Model 2701 Instrument Networking Instruction Manual. The following
example demonstrates correct settings for the Model 2701 and the PC:
Example – Assume the 2701 is using the default IP Address and Subnet mask:
Model 2701 settings:
IP Address = 192.168.000.002
192.168.000 is the Network ID
002 is the Node Designation
Subnet mask = 255.255.255.000
PC settings:
IP Address = 192.168.000.004
192.168.000 is the Network ID
004 is the Node Designator
Subnet mask = 255.255.255.000
As shown above for the IP Address, the Network ID of the 2701 and PC must be the same.
The Node Designators must be different. For the Subnet, mask must be the same for the
2701 and the PC.