10-30 Remote Operations Model 2701 User’s Manual
Terminator
The Model 2701 can be configured to terminate each program message that it transmits to
the controller with any of the following combinations of <CR> and <LF>.
<CR> Carriage return
<CR+LF> Carriage return and line feed
<LF> Line feed
<LF+CR> Line feed and carriage return
Selecting and configuring RS-232 interface
After selecting (enabling) the RS-232 interface, you will then set the baud rate, flow
control, and terminator.
1. Press the SHIFT key and then the RS-232 key. The RS 232 ON or RS 232 OFF
message will be displayed.
2. If the RS-232 is already ON, press ENTER and proceed to step 3. Otherwise, press
the key to place the cursor on OFF, press the or key to display the ON state,
and then press ENTER.
NOTE Enabling (ON) the RS-232 interface disables (OFF) the Ethernet. Disabling the
RS-232 interface enables the Ethernet. Changing the state reboots the
Model 2701.
3. To retain the presently displayed BAUD rate, press ENTER and proceed to step 4.
Otherwise, press the key to place the cursor on the baud rate value, use the or
key to display the desired baud rate, and then press ENTER.
4. To retain the presently displayed FLOW control, press ENTER and proceed to
step 5. Otherwise, press the key to place the cursor on the flow control setting,
use the or key to display the alternate flow control setting, and then press
ENTER.
5. To retain the presently displayed terminator (Tx TERM), press ENTER. Otherwise,
press the key to place the cursor on the terminator setting, use the or key to
display the desired terminator, and then press ENTER.