Model 2701 User’s Manual Status Structure 11-9
The bits of the Status Byte Register are described as follows:
Bit B0, Measurement Summary Bit (MSB) — Set summary bit indicates that an
enabled measurement event has occurred.
Bit B1 — Not used.
Bit B2, Error Available (EAV) — Set summary bit indicates that an error or status
message is present in the Error Queue.
Bit B3, Questionable Summary Bit (QSB) — Set summary bit indicates that an
enabled questionable event has occurred.
Bit B4, Message Available (MAV) — Set summary bit indicates that a response
message is present in the Output Queue.
Bit B5, Event Summary Bit (ESB) — Set summary bit indicates that an enabled
standard event has occurred.
Bit B6, Master Summary Status (MSS) — Set bit indicates that an enabled
summary bit of the Status Byte Register is set.
Bit B7, Operation Summary (OSB) — Set summary bit indicates that an enabled
operation event has occurred.
Service request enable register
The generation of a service request is controlled by the Service Request Enable Register.
This register is programmed by you and is used to enable or disable the setting of bit B6
(MSS) by the Status Summary Message bits (B0, B2, B3, B4, B5, and B7) of the Status
Byte Register. As shown in Figure 11-3, the summary bits are logically ANDed (&) with
the corresponding enable bits of the Service Request Enable Register. When a set (1)
summary bit is ANDed with an enabled (1) bit of the enable register, the logic “1” output
is applied to the input of the OR gate and, therefore, sets the MSS bit in the Status Byte
Register.
The individual bits of the Service Request Enable Register can be set or cleared by using
the *SRE common command. To read the Service Request Enable Register, use the *SRE?
query command. The Service Request Enable Register clears when power is cycled or a
parameter value of 0 is sent with the *SRE command (*SRE 0). The commands to
program and read the SRQ Enable Register are listed in Table 11-3.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
BrandKeithley
Model2701
CategoryMultimeter
LanguageEnglish

Related product manuals