Model 2701 User’s Manual Triggering 8-3
Control source and event detection
The control source holds up operation until the programmed event occurs and is detected.
The control sources are described as follows:
Immediate — With this control source, event detection is immediately satisfied
allowing operation to continue.
External — Event detection is satisfied for any of the following three conditions:
An input trigger via the Trigger Link line EXT TRIG is received.
The front panel TRIG key is pressed. (The Model 2701 must be taken out of
remote before it will respond to the TRIG key. Use the LOCAL key or send
SYSTem:LOCal over the Ethernet or RS-232.)
Trigger command (*TRG) is received over the Ethernet or RS-232.
Delay (auto or manual)
A programmable delay is available after event detection. It can be set manually or an auto
delay can be used. With auto delay selected, the instrument automatically selects a delay
period that will provide sufficient settling for function changes, autorange changes, and
multi-phase measurements.
Normal measurement stateWith auto delay selected and the External or Bus control
source selected, the Model 2701 selects a delay based on the selected voltage range. The
auto delay period cannot be adjusted by the user. The auto delays are listed in Table 8-1.
With one of the other control sources selected, the auto delay is 0.000s for all functions
and ranges.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
BrandKeithley
Model2701
CategoryMultimeter
LanguageEnglish

Related product manuals