1-30 Getting Started Model 2701 User’s Manual
Basic DMM measurements — front panel inputs
NOTE See Section 3 for details on basic DMM operation.
The Model 2701 is shipped from the factory to power-up to factory defaults. The instru-
ment powers up to a setup that continuously measures DC volts. Some of the default set-
tings for the DCV function include auto range enabled, 6½-digit resolution, filter enabled,
and slow reading rate. These settings provide a good starting point and in many cases, do
not need to be changed.
“Starting-point” default settings are also provided for the other measurement functions.
Therefore, to perform basic measurements, simply select the desired function, and
“tweak” the setup (range, rate, filter, digits, etc.) as required.
For remote programming, the instrument is typically used in a non-continuous measure-
ment mode. In this mode, the user (via remote command programming) specifies the num-
ber of measurements to perform. *RST defaults place the instrument in a non-continuous
measurement mode. Most of the other settings for factory and *RST defaults are the same.
For remote programming, the following command is used to select function.
NOTE Items in brackets ([]) are optional and do not need to be included. Upper case
characters are required. Lower case characters are optional and need not be
included.
[SENSe[1]]:FUNCtion <func> ’Select measurement function.’
<func> =’VOLTage[:DC]’ DCV
’VOLTage:AC’ ACV
’CURRent[:DC]’ DCI
’CURRent:AC’ ACI
’RESistance’ 2
’FRESistance’ 4
’FREQuency’ FREQ
’PERiod’ PERIOD
’TEMPerature’ TEMP
Each function can have its own unique setup configuration (i.e., range, digits, speed, etc.).
For example, the following command words select range and digits:
RANGe[:UPPer] <n> ’ Specify expected reading.
RANGe:AUTO <b> ’ Enable (ON) or disable (OFF) auto range.
DIGits ’ Set display resolution; 3.5, 4.5, 5.5, or 6.5 (digits).
The following examples demonstrate how to include the function name in the command
string for configuration commands.
VOLT:RANG 10 ’ Select 10V range for DCV.
RES:RANG:AUTO ON ’ Enable auto range for 2.
CURR:DIG 4.5 ’ Set DCI for 4½-digit resolution.
NOTE See Section 4 for details on setting range, digits, rate, bandwidth, and filter.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
BrandKeithley
Model2701
CategoryMultimeter
LanguageEnglish

Related product manuals