EasyManuals Logo
Home>Keithley>Multimeter>2701

Keithley 2701 User Manual

Keithley 2701
506 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #270 background imageLoading...
Page #270 background image
8-6 Triggering Model 2701 User’s Manual
Reading hold (autosettle)
With hold enabled (HOLD annunciator on), the first processed reading becomes the
“seed” reading and operation loops back within the device action block. After the next
reading is processed, it is checked to see if it is within the selected hold window (0.01%,
0.1%, 1%, or 10%) of the “seed” reading. If the reading is within the window, operation
again loops back within the device action block. This looping continues until the specified
number (2 to 100) of consecutive readings are within the window. If one of the readings is
not within the window, the instrument acquires a new “seed” reading and the hold process
continues.
When a hold reading is acquired, an audible beep is sounded (if enabled) and the reading
is considered a “true measurement.” The reading is held on the display until an “out of
window” reading occurs to restart the hold process.
For remote operation, the hold process seeks a new “seed” once it has been satisfied and
the reading has been released. For basic front panel operation, the hold process does not
seek a new “seed” until the held condition is removed.
NOTE Hold cannot be used when scanning.
Hold example
1. Press SHIFT and then HOLD to display the present window (0.01%, 0.1%, 1%, or
10%).
2. To change the window, press the
or key to display the desired window.
3. Press ENTER. The present hold count is displayed (2 to 100).
4. To change the hold count, use the
, , , and keys to display the desired count.
5. Press DCV to measure DC voltage.
6. Apply the test signal to the input of the Model 2701. Once the signal becomes stable
enough to satisfy the hold condition, the reading is released and the beeper sounds (if
enabled).
7. Remove the hold condition by disconnecting the signal from the input. Hold will then
seek a new “seed.
8. To disable HOLD, press SHIFT and then HOLD.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
BrandKeithley
Model2701
CategoryMultimeter
LanguageEnglish

Related product manuals