G-4 KE2700 Instrument Driver Examples Model 2701 User’s Manual
Name Manual Reference Brief Description
Advance4 None Use Case 4 — Two scans using 7708 module:
40 channel DCV scan (1V range). Configuration saved in
User Setup 1.
20 channel 4 scan. Configuration saved in User Setup 2.
Models 2700 and 2701 – 100 range.
Model 2750 – 10 range, dry-circuit ohms enabled.
Setup 1 or Setup 2 recalled to perform scan.
Measurement speed (rate) – 0.1 plc.
DCV input divider – Enabled (10M input impedance).
Filter – Disabled (no filtering).
Buffer – Store 40 DCV reading strings, 20 4 reading strings.
Buffer elements include reading only.
Limits (DCV scan) – Limit 1 (all channels) = 20mV, Master
Latch enabled.
Triggering – Bus control source.
Data retrieval – SRQ if limit fails.
Advance5 None Use Case 5 — 32-channel scan using 7701 module:
Common-side 4-wire ohms measurements (CSIDe mode).
Dry-circuit ohms option for Model 2750.
Install jumpers to connect Input Hi and Sense Hi directly to
DUT (common-side bus).
Install jumpers to connect channel 35 to Sense Lo and Input
Lo.
Buffer – Store 32 reading strings. Buffer elements include
reading only.
Triggering – Immediate control source.
Data retrieval – SRQ when buffer full.
Table G-1 (continued)
Visual Basic and CVI (C) examples

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
BrandKeithley
Model2701
CategoryMultimeter
LanguageEnglish

Related product manuals