4-4 Range, Digits, Rate, Bandwidth, and Filter Model 2701 User’s Manual
Remote programming range
Range commands
The commands to set range are listed in Table 4-2. Additional information on these
commands follow the table.
NOTE Query commands and some optional command words are not included in
Table 4-2. All commands for the SENSe subsystem are provided in Table 15-5.
Table 4-2
Range commands
Commands
1, 2
Description Default
[SENSe[1]] Optional root command.
:VOLTage[:DC]:RANGe <n> [, <clist>] Select DCV range; <n> = 0 to 1010 (V). 1000
:VOLTage[:DC]:RANGe:AUTO <b> [,
<clist>]
Control DCV auto range; <b> = ON or OFF. ON
:VOLTage:AC:RANGe <n> [, <clist>] Select ACV range; <n> = 0 to 757.5 (V). 750
:VOLTage:AC:RANGe:AUTO <b> [, <clist>] Control ACV auto range; <b> = ON or OFF. ON
:CURRent[:DC]:RANGe <n> [, <clist>] Select DCI range; <n> = 0 to 3 (A). 3
:CURRent[:DC]:RANGe:AUTO <b> [,
<clist>]
Control DCI auto range; <b> = ON or OFF. ON
:CURRent:AC:RANGe <n> [, <clist>] Select ACI range; <n> = 0 to 3 (A). 3
:CURRent:AC:RANGe:AUTO <b> [,
<clist>]
Control ACI auto range; <b> = ON or OFF. ON
:RESistance:RANGe <n> [, <clist>] Select 2 range; <n> = 0 to 120e6 (). 120e6
:RESistance:RANGe:AUTO <b> [, <clist>] Control 2 auto range; <b> = ON or OFF. ON
:FRESistance:RANGe <n> [, <clist>] Select 4 range; <n> = 0 to 120e6 (). 120e6
:FRESistance:RANGe:AUTO <b> [, <clist>] Control 4 auto range; <b> = ON or OFF. ON
Channel list parameter:
<clist> = (@SCH)
where: S = Mainframe slot number (1 or 2)
CH = Switching module channel number (must be 2 digits)
Examples: (@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
1. The <clist> parameter is used to configure one or more channels for a scan. Each channel in the <clist> must be set to the function
specified by the range command. If not, a conflict error (-221) will occur. For example, VOLTage:AC:RANGe 1, (@101) is only
valid if scan channel 101 is set for the ACV function.
2. [:DC] is optional for the commands to set DCV and DCI range.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
BrandKeithley
Model2701
CategoryMultimeter
LanguageEnglish

Related product manuals