Model 2701 User’s Manual SCPI Reference Tables 15-19
Table 15-6
STATus command summary
Command Description
Default
parameter Ref SCPI
STATus (Note 1) Sec 11 ✓
:MEASurement Measurement event registers:
[:EVENt]? Read the event register.
6
(Note 2)
:ENABle <NDN> or <NRf> Program the enable register. (Note 3)
:ENABle? Read the enable register.
6
:CONDition? Read the condition register.
6
:OPERation Operation status registers: ✓
[:EVENt]? Read the event register.
6
(Note 2) ✓
:ENABle <NDN> or <NRf> Program the enable register. (Note 3) ✓
:ENABle? Read the enable register.
6
✓
:CONDition? Read the condition register.
6
✓
:QUEStionable Questionable status registers: ✓
[:EVENt]? Read the event register.
6
(Note 2) ✓
:ENABle <NDN> or <NRf> Program the enable register. (Note 3) ✓
:ENABle? Read the enable register.
6
✓
:CONDition? Read the condition register.
6
✓
:PRESet Return status registers to default states. ✓
:QUEue Path to access error queue: ✓
[:NEXT]? Read the most recent error message. (Note 4) ✓
:ENABle <list> Specify error and status messages for
queue.
(Note 5) ✓
:ENABle? Read the enabled messages. ✓
:DISable <list> Specify messages not to be placed in
queue.
(Note 5)
:DISable? Read the disabled messages.
:CLEar Clears all messages from Error Queue.
Notes:
1. Commands in this subsystem are not affected by *RST and SYSTem:PRESet. The effects of cycling power, *CLS, and
STATus:PRESet are explained by the following notes.
2. Event Registers: Power-up and *CLS – Clears all bits of the registers.
STATus:PRESet – No effect.
3. Enable Registers: Power-up and STATus:PRESet – Clears all bits of the registers.
*CLS – No effect.
4. Error Queue: Power-up and *CLS – Clears the Error Queue.
STATus:PRESet – No effect.
5. Enable/Disable Error Queue Messages: Power-up – Clears list of messages.
*CLS and STATus:PRESet – No effect.
6. The format for the response message (decimal, hexadecimal, octal, or binary) is set by the FORMat:SREGister command.