Model 2701 User’s Manual SCPI Signal Oriented Commands 13-5
Query CONFigure? Query the selected function.
Description <clist> included — When the <clist> parameter is included with
CONFigure command, the specified channel(s) for the scanlist assumes
the *RST default settings for the specified function. Range can also be
set for the channel(s) by including the <rang> parameter. If the
resolution parameter (<res>) is included, it will be ignored.
The present measurement function and the trigger model settings are not
affected when the CONFigure command is sent with the <clist>
parameter.
<clist> not includedWhen the <clist> parameter is not included, the
CONFigure command configures the instrument for subsequent
measurements on the specified function. Range and resolution can also
be set for the specified function.
This command places the instrument in a “one-shot” measurement
mode. You can then use the READ? command to trigger a measurement
and acquire a reading (see READ?).
When this command is sent without the <clist> parameter, the
Model 2701 will be configured as follows:
The function specified by this command is selected. If specified,
range and/or resolution are also set.
All controls related to the selected function are defaulted to the
*RST values.
Continuous initiation is disabled (INITiate:CONTinuous OFF).
The control source of the Trigger Model is set to Immediate.
The count values of the Trigger Model are set to one.
The delay of the Trigger Model is set to zero.
The Model 2701 is placed in the idle state.
All math calculations are disabled.
Buffer operation is disabled. A storage operation presently in
process will be aborted.
Autozero is enabled.
Programming examples
Programming example #1 — The following command configures
scanlist channels 101 through 105 for 4-wire resistance measurements
on the 1M range.
CONF:FRES 1e6, (@101:105)
Programming example #2 — The following command selects the
DCV function, 10V range, 3½ digit resolution, and performs the “no
<clist>” CONFigure operations:
CONF:VOLT 10, 3.5

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
BrandKeithley
Model2701
CategoryMultimeter
LanguageEnglish

Related product manuals