9-10 Sweep Operation
Testing processThe test uses seven SMLs (source memory locations). However, only
four memory locations are used for each tested diode. If the diode is installed correctly, tests at
locations 001, 002, 003, and 004 are performed. If the diode is installed backwards, tests at
locations 001, 005, 006, and 007 are performed. To sweep four memory locations, the sweep
count must be set to four. The source memory sweep is summarized as follows:
SML 001 — Compliance Test
Limit 1 test – Fail if in compliance, branch to source memory location 005 for “pass”
condition.
Summary – Limit 1 test is configured such that if the diode is installed correctly in the
test fixture, it will fail the compliance test and operation will proceed to the tests at
memory locations 002, 003, and 004. If the diode is installed backwards, it will pass the
compliance test, and operation will branch around locations 002, 003, and 004 to per-
form the tests at locations 005, 006, and 007. Source Memory Location 002 - Forward
Voltage Test (diode installed correctly).
SML 002 — Forward Voltage Test
Source I, Measure V.
Limit 2 test – Min/max limits for voltage reading.
Summary – The voltage measurement and the result of the test (pass or fail) is stored in
the buffer.
SML 003 — Reverse Breakdown Test
Source -I, Measure V.
Limit 2 test – Min/max limits for voltage reading.
Summary – The voltage measurement and the result of the test (pass or fail) is stored in
the buffer.
SML 004 — Leakage Current Test
Source -V, Measure I.
Limit 2 test – Min/max limits for current reading.
Summary – The current measurement and the result of the test (pass or fail) is stored in
the buffer.
SML 005 — Forward Voltage Test
Source -I, Measure V.
Limit 2 test – Min/max limits for voltage reading.
Summary – This test is the same as the test at memory location 002, except the source
current is reversed to properly bias the diode that was installed backwards.
SML 006 — Reverse Breakdown Test
Source +I, Measure V.
Limit 2 test – Min/max limits for voltage reading.
Summary – This test is the same as the test at memory location 003, except the source
current is reversed to properly bias the diode that was installed backwards.

Table of Contents

Other manuals for Keithley 6430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6430 and is the answer not in the manual?

Keithley 6430 Specifications

General IconGeneral
BrandKeithley
Model6430
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals