Triggering 10-11
Step 6: Set trigger out events to MEAS=ON (all others to OFF)
Select EVENTS, then press ENTER. Select MEAS=OFF and toggle the value to ON using
the
and keys. Press ENTER, and then press EXIT to return to the CONFIGURE TRIG-
LAYER menu.
Step 7: Set trigger count to 10
Use the right cursor key to scroll to the far right of the menu selections and select COUNT,
then press ENTER. Set the count to 10, then press the ENTER key. Press the EXIT key twice to
leave CONFIGURE TRIGGER menus.
Step 8: Enable auto output off
Press the CONFIG key and then the ON/OFF key to access the CONFIGURE OUTPUT
menu. Select AUTO-OFF, then press ENTER. Select ENABLE, then press ENTER. Select
ALWAYS, then press ENTER. Press the EXIT key to leave the CONFIGURE OUTPUT menu.
Switching mainframe setup
Step 1: Restore bench defaults
Press the MENU key, select SAVESETUP, and then press ENTER. From the SAVESETUP
menu, select RESET, then press ENTER. Press ENTER to confirm the action. Press ENTER to
return to the SETUP MENU. Press EXIT to leave the SETUP MENU. Press EXIT to leave the
MAIN MENU.
Step 2: Set up scan list: 1!1 - 1!10
Press the SCAN LIST key. Press 1, 1 - 1, 1, 0 , then press the ENTER key.
Step 3: Set then number of scans to 1
Press the SCAN key, select SCAN-CONTROL, and then press ENTER. From the SCAN
CONTROL menu, select NUMBER-OF-SCANS, then press ENTER. Select ENTER-SCAN-
COUNT, then press ENTER. Set the count to 1, then press ENTER. Press the EXIT key to
return to the CONFIGURE SCAN menu.
Step 4: Set channel spacing to trigger-link
Select CHAN-CONTROL from the CONFIGURE SCAN menu, then press ENTER. Select
CHANNEL-SPACING from the CHANNEL CONTROL menu, then press ENTER. Select
TRIGLINK, then press ENTER. Select ASYNCHRONOUS, then press ENTER. Press
ENTER, ENTER, then EXIT, EXIT, EXIT to leave the CONFIGURE SCAN menu.

Table of Contents

Other manuals for Keithley 6430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6430 and is the answer not in the manual?

Keithley 6430 Specifications

General IconGeneral
BrandKeithley
Model6430
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals