EasyManuals Logo
Home>Keithley>Measuring Instruments>6430

Keithley 6430 User Manual

Keithley 6430
558 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #373 background imageLoading...
Page #373 background image
SCPI Command Reference 17-63
:ADVanced:NTOLerance <NRf>
[:SENSe[1]]:AVERage:ADVanced:NTOLerance <NRf> Set moving lter noise lter
Parameters <NRf> = 0 to 105 Specify lter noise lter in %
Query :NTOLerance? Query lter noise lter value
Description When the advanced lter is enabled, using the next command, a noise win-
dow is used with the moving lter. This command is used to specify the
noise window. If readings are within the noise window, the moving lter
operates normally. If, however, a reading falls outside the window, the
buffer is ushed of old readings and lled with the new reading.
:ADVanced[:STATe] <b>
[:SENSe[1]]:AVERage:ADVanced[:STATe] <b> Enable/disable advanced lter
Parameters <b> = 0 or OFF Disable advanced lter
1 or ON Enable advanced lter
Query [:STATe]? Query state of advanced lter
Description This command is used to enable or disable the advanced lter. When
enabled, the noise window is used with the moving lter. When disabled,
the noise window is not used.

Table of Contents

Other manuals for Keithley 6430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6430 and is the answer not in the manual?

Keithley 6430 Specifications

General IconGeneral
BrandKeithley
Model6430
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals