A-2 Specifications
SOURCE SPECIFICATIONS
1
Voltage Programming Accuracy (4-wire sense)
2
Accuracy (1 Year) Noise
Programming 23°C ±5°C (peak-peak)
Range Resolution ±% rdg. + volts 0.1Hz – 10Hz
200.000 mV 5 µV 0.02% + 600 µV 5 µV
2.00000 V 50 µV 0.02% + 600 µV 50 µV
20.0000 V 500 µV 0.02% + 2.4 mV 500 µV
200.000 V 5 mV 0.02% + 24 mV 5 mV
TEMPERATURE COEFFICIENT (0°–18°C & 28°–40°C): ±(0.15 × accuracy specification)/°C.
MAX. OUTPUT POWER: 2.2W (four quadrant source or sink operation).
SOURCE/SINK LIMITS: ±21V @ ±105mA, ±210V @ ±10.5mA.
VOLTAGE REGULATION: Line: 0.01% of range. Load: 0.01% of range + 100µV.
NOISE 10Hz1MHz (p-p): 10mV.
OVER VOLTAGE PROTECTION: User selectable values, 5% tolerance. Factory default = None.
CURRENT LIMIT: Bipolar current limit (compliance) set with single value. Min. 0.1% of range.
Current Programming Accuracy (with remote preamp)
Accuracy (1 Year)
1
Noise
Programming 23°C ±5°C (peak-peak)
Range Resolution ±% rdg. +amps 0.1Hz 10Hz
1.00000 pA 50 aA 1.0 % + 10 fA 5 fA
10.0000 pA 500 aA 0.50 % + 30 fA 10 fA
100.000 pA 5 fA 0.15 % + 40 fA 20 fA
1.00000 nA 50 fA 0.050 % + 200 f A 50 fA
10.0000 nA 500 fA 0.050 % + 2 pA 500 fA
100.000 nA 5 pA 0.050 % + 20 pA 3 pA
1.00000 µA 50 pA 0.050 % + 300 pA 20 pA
10.0000 µA 500 pA 0.050 % + 2 nA 200 pA
100.000 µA 5 nA 0.031 % + 20 nA 500 pA
1.00000 mA 50 nA 0.034 % + 200 nA 5 nA
10.0000 mA 500 nA 0.045 % + 2 µA 50 nA
100.000 mA 5 µA 0.066 % + 20 µA 500 nA
Current Programming Accuracy (without remote preamp)
Accuracy (1 Year)
1
Noise
Programming 23°C ±5°C (peak-peak)
Range Resolution ±% rdg. +amps 0.1Hz 10Hz
1.00000 µA 50 pA 0.035 % + 600 pA 20 pA
10.0000 µA 500 pA 0.033 % + 2 nA 200 pA
100.000 µA 5 nA 0.031 % + 20 nA 500 pA
1.00000 mA 50 nA 0.034 % + 200 nA 5 nA
10.0000 mA 500 nA 0.045 % + 2 µA 50 nA
100.000 mA 5 µA 0.066 % + 20 µA 50 nA
TEMPERATURE COEFFICIENT (0°–18°C & 28°–40°C): ±(0.15 × accuracy specification)/°C.
MAX. OUTPUT POWER: 2.2W (four quadrant source or sink operation).
SOURCE/SINK LIMITS: ±10.5mA @ 210V, ±105mA @ 21V.
CURRENT REGULATION: Line: 0.01% of range. Load: 0.01% of range + 1fA.
VOLTAGE LIMIT: Bipolar voltage limit (compliance) set with single value. Min. 0.1% of range.
1
For sink mode, 1pA to 100mA range, accuracy is ±(0.15% + offset
*
4).
2
Voltage source accuracies are not affected by the remote preamp.

Table of Contents

Other manuals for Keithley 6430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6430 and is the answer not in the manual?

Keithley 6430 Specifications

General IconGeneral
BrandKeithley
Model6430
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals