EasyManuals Logo
Home>Mindray>Diagnostic Equipment>TEX20

Mindray TEX20 Service Manual

Mindray TEX20
445 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #150 background imageLoading...
Page #150 background image
Revision:1.0(2023-01-12)
151
Table 5–2 Tools and measuring devices(continued)
Consumables Quantity
Remarks
Mild soapy water About 400 mL /
Dry soft cloth or cotton cloth About 5 pcs /
55..11..11..33 Routine Maintenance Items
Table 5–3 Maintenance items and maintenance frequency
SN Item Recommend-
ed
Frequency
Operator Method
Terminal
users
Engineer
1 Cleaning the Dust-
Proof Filter
Once per
month
For
details,
see
5.1.2.1
Flow of
Cleaning
2 Clean the touch
screen.
Once per
month
3 Clean the probe
(acoustic head
part).
After each use
4
Clean the probe
cable and
connector housing.
Once per
month
5
Cleaning the
Holders
Once per
month
6 Clean the cover Once per
month
7 Peripherals to Be
Cleaned
Once per
month
8 Check the probe
surface.
1 time/day
For
details,
see
5.1.3.1
General
Inspections
9 Check the power
cord and plug.
Once per
month
10 Check the battery. Once per 3–6
months
11 Check the functions
of peripherals and
optional
accessories.
Once per year
12 Mechanical Safety
Inspection
Once per year
Diagnostic Ultrasound System
Service Manual
5 Maintenance

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish